You are viewing the site in preview mode

Skip to main content

Table 1 Summary of imaging parameters for phantom and in vivo studies*

From: Rapid phase-modulated water-excitation steady-state free precession for fat-suppressed cine cardiovascular MR

Sequences Interpulse Phase Evolution (°) Interpulse Delay (ms) Total Pulse Duration (ms) TR for Phantom Studies (ms) TR for in vivo Studies (ms) Component Pulse Flip Angles (°) Resultant Flip Angle (°)
Standard SSFP NA NA 0.6 9.68 3.1 70 70
1-2-1 WE-SSFP 180 2.2 5.0 9.68 8.9 17.7 – 35.4 – 17.7 70
1-1 WE-SSFP 180 2.2 2.8 9.68 6.5 35.4 – 35.4 70
1-1 WE-SSFP 90 1.1 1.7 9.68 4.0 56.4 – 56.4 70
  1. * NA: not applicable